SuSix accessories

Find the right accessories for your SuSix equipment.
155276 Sensor immersion bundle kit for SuSix® sensor ISO, consisting of:
- 155240 SuSix® sensor adaptor for immersion ISO
- 155260 PVC pipe 50mm - PN10 (ø50 x 2,4mm) (Supplied local)
- 155268 PVC CAP ISO 50mm w. PG9 Cable Glands
- 579055 PVC glue bottle with 120 gram
- 200205 Universal bracket
- 200210 Wall support for universal bracket
- 200215 Extension for universal bracket
155277 Sensor immersion bundle kit for SuSix® sensor NPT, consisting of:
- 155241 SuSix® sensor adaptor for immersion NPT
- 155261 PVC pipe 2" (ø1.9"x0.1") (Supplied local)
- 155269 PVC CAP US w. PG9 Cable Glands
- 579055 PVC glue bottle with 120 gram
- 200205 Universal bracket
- 200210 Wall support for universal bracket
- 200215 Extension for universal bracket
SuSix sensor Wiper maintenance
Maintenance interval - Every Two Years
SuSix® sensors with wipers call for replacement of the wiper drive shaft gasket every 2 years.
Reservations
Replacement of wiper and gasket must be carried out by MJK authorized and trained personnel in an electrostatic discharge (ESD) secured area.
206350 Wiper maintenance kit
Use Wiper replacement of MJK SuSix® Sensor (part no. 206312) with cleaning system/wiper from serial no. 51500 and higher.
The wiper replacement part 206350 includes gasket.
Contents
Part no. 206350 including:
• 1 pc. screwdriver
• 1 pc. wiper mounted on shaft with two O-rings
206355 Wiper and gasket maintenance kit
Wiper and gasket replacement of MJK SuSix® Sensor (part no. 206312) with cleaning system/wiper from serial no. 49482 to 51499 incl.
Contents
Part no. 206355 including:
• 1 pc. screwdriver
• 7 pcs. screws
• 5 pcs. wipers
• 5 pcs. gaskets
Required accessories
206360 Susix® Gasket tool
206356 Wiper maintenance kit
- 206355 Wiper maintenance kit
Wiper replacement of MJK SuSix® Sensor (part no. 206312) with cleaning system/wiper from serial no. 49482 to 51499 incl.
Contents
Part no. 206356 including:
• 1 pc. screwdriver
• 7 pcs. screws
• 5 pcs. wipers
206360 Susix® Gasket tool
206370 Susix® Test Unit for Turbidity
Use
For test and check of the factory set calibration for turbidity of SuSix® sensors. SuSix® sensors holds a fixed factory calibration for turbidity. The Turbidity Calibration Test Unit is therefore a tool for check of calibration and not a tool for recalibration of SuSix® sensors.
The Test Unit simulates a turbidity close to pure water. A SuSix® sensor will give rise to a reading of approx. 1 NTU / FTU when challenged to a Test Unit.
MJK Field Linkâ„¢ 840110
For support of MagFlux, SuSix, Oxix, mA/Bus Converter and Connect / Mµ Connect.
Features and functions
MJK Field Linkâ„¢ is used for communication between your MJK instrument to a PC by use of a standard USB cable.
Log Data
Import and show log Data and export as CSV files.
Firmware
Upload latest firmware using MJK Field Linkâ„¢ to Display and Converter
System requirements
- Microsoft Windows XP sp2/ Vista /Win 7, 32bit
- Java
- USB 2.0 connection, Use with USB cable 691095
691095 USB Cable
PC Mini USB/USB Cable
691075 Communication cable
- 2x2x0.5mm2 (AWG24) Shielded twistet pair
- Colours: Red, white, black, Grey
- PUR
Communication cable for serial connection between MJK instruments, designed for RS232 and Modbus communication.
The cable is protected by a thick PUR protection so it can withstand the tension for outdoor mounting.